| Customization: | Available |
|---|---|
| Type: | Defibrillator |
| Certification: | ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Specifications
| Two-button operation Biphasic waveform energy output Three-step defibrillation process Extensive audible and visual prompts for the operator Lock-out protection to prevent inadvertent defibrillation Non-rechargeable Battery: 12V DC 2.8Ah Capacity: 100 discharges at 200J or 120 discharges at 150J |
| Physical characteris
|
Dimensions | 303*216*89mm |
| Weight | 2.0KG | |
| Operating Temperature | 0°C to 40°C | |
| Operating Humidity | Relative humidity between 30% and 95% (non-condensing) | |
| Storage Temperature (without battery) | -20°C to 55°C | |
|
Storage Humidity (without battery) |
Up to 95% (non-condensing) |
|
|
Defibrillator
|
Output |
Biphasic Truncated Exponential |
|
Energy Sequence |
150, 150, 200J |
|
|
Charge Time |
8 sec. to 150J; 10 sec. to 200J (New Battery) |
|
|
Analysis Time |
9 sec. |
|
|
The maximum time from the initiation of rhythm analysis to readiness for discharge with a new battery |
Less than 30 seconds |
|
|
The maximum time from the initiation of rhythm analysis to readiness for discharge after 6 shocks |
Less than 35 seconds |
|
|
The maximum time from initially switching power on to readiness for discharge |
Less than 25 seconds |
|
|
Audible Prompts |
22 audible prompts |
|
|
Visual Prompts |
LED prompts |
|
|
Controls |
Two buttons-ON/Off, Shock |
|
|
Output Energy Accuracy |
±15% into any impedance from 25 to 175Ω |
|
|
The Maximum Voltage |
1200±50V |
|
|
Output disabled when the patient impedance is outside limits |
20Ωto 200Ω |
|
|
Battery
|
Part No. |
CR123A-4*2 |
|
Non-rechargeable |
12V DC, 2.8Ah |
|
|
Capacity |
At 150 Joules |